Melamine Powder -Irradiated Formula: C3N6H6 Core Loss

Submitted by Susana Trasobares, November 19, 2007.

Source / Purity: Aldrich 99%

Author Comments: The experiment was performed with the chronospectroscopy mode, every spectrun was acquired using an acquisition time of 9 millisecond. The submitted spectrum is a collection of 10 spectra. Melamine is really sensitive to beam damage, after 0.4 seconds the origianl melamine spectra is transformed into the spectrun submitted here. Analyst: Susana Trasobares. Temperature: Room.

Register / log in to set favourite spectra and overlay plots.
loading plot..

You can manually set the axis limits.

Spectrum Metadata

Specimen Name
VG HB501
Spectrum Type
Core Loss
Specimen Formula
C3N6H6
Data Range
244.41 eV - 457.19 eV
Source and Purity
Aldrich 99%
Elemental Edges
C_K
Keywords
imported from old site
Alternative URL
Old EELS DB
Microscope Name / Model
VG HB501
Gun Type
Cold FEG
Incident Beam Energy
100 kV
Monochromated
No
Dispersion
0.208 eV/pixel
Convergence Semi-angle
15 mrad
Collection Semi-angle
24 mrad
Probe Size
0.5 nm2
Detector
Parallel: 666 Gatan
Dark Current Correction
Yes
Gain Variation Spectrum
Yes
Reference
Susana Trasobares, Christine Kolczewski, Riitta Räty, Niclas Borglund, Arianna Bassan, Gilles Hug, Christian Colliex, Stefan Csillag, Lars G. M. Pettersson (2003). Monitoring the Decomposition of Melamine in the Solid Phase by Electron Energy Loss Chronospectroscopy J. Phys. Chem. A 107:2 228-235