Vanadium Oxide Formula: V2O5 Low Loss

Submitted by Ferdinand Hofer, January 15, 2008.

Source / Purity: Dangsheng Su, Fitz-Haber-Institute, Berlin

Author Comments: Not calibrated Analyst: Gerald Kothleitner. Temperature: Room.

Register / log in to set favourite spectra and overlay plots.
loading plot..

Chart x axis limited to > 2 eV due to low loss spectrum type. Click here to remove this limit. You may also manually set the axis limits.

Spectrum Metadata

Specimen Name
Tecnai F20 monochromator
Spectrum Type
Low Loss
Specimen Formula
V2O5
Data Range
-2 eV - 18.46 eV
Source and Purity
Dangsheng Su, Fitz-Haber-Institute, Berlin
Keywords
imported from old site
Associated Spectra
CL vanadium oxide
Alternative URL
Old EELS DB
Microscope Name / Model
Tecnai F20 monochromator
Gun Type
feg Schottky
Incident Beam Energy
200 kV
Resolution
0.2 eV
Monochromated
Yes
Dispersion
0.02 eV/pixel
Convergence Semi-angle
2 mrad
Collection Semi-angle
7.5 mrad
Detector
GIF: GIF Tridiem
Dark Current Correction
Yes
Gain Variation Spectrum
Yes
Reference
D.S.Su, H.W.Zandbergen, P.C.Tiemeijer, G. Kothelitner et al., Micron 34, 235-238, 2003